The KMP system brings indirect luminous intensity distribution (LID) measurement into compact environments. With its lightweight and space-efficient design, KMP can be perfectly integrated into production lines or inside smaller laboratories without major infrastructure changes.

KMP employs a calibrated LMK imaging photometer to deliver photometric outputs including luminous intensity (cd), illuminance (lx), and color coordinates (e.g. xy, uv). For larger measurement objects, the BV Room variant complements the product line.

Modular Components and Calibration Options

How KMP Works

Your device under test illuminates a precisely calibrated reflective screen and the LMK imaging photometer records the reflected luminance distribution. From the captured image, our software computes luminous intensity distributions and illuminance and derived outputs such as partial luminous flux.

 
Basic technical data Automotive Airfield
Size of the device under test
< 120 mm < 190 mm 1
Typical measuring distance 25 m – ∞  25 m – ∞ 
Object field horizontal ±24°, vertical ±15°
horizontal ±22°, vertical ±15°
Resolution ≥ 0.01° ≥ 0.1°
Measuring range 0.3 cd – 50 Mcd 5 cd – 10 Mcd
Measurable contrast2 1:10 000 1:100
Measuring time seconds seconds
Dimensions 980 mm × 850 mm × 650 mm (L × W × H) 980 mm × 850 mm × 650 mm (L × W × H)
Basic technical data
Size of the device under test Automotive: < 120 mm
Airfield: < 190 mm 1
Typical measuring distance Automotive: 25 m – ∞ 
Airfield: 25 m – ∞ 
Object field Automotive: hor. ±24° by vert. ±15° 
Airfield: Hor. ±22°; vert. ±15°
Resolution Automotive: ≥ 0.01°
Airfield: ≥ 0.1°
Measuring range Automotive: 0.3 cd – 50 Mcd
Airfield: 5 cd – 10 Mcd
Measurable contrast2 Automotive: 1:10 000
Airfield: 1:100
Measuring time Automotive: seconds
Airfield: seconds
Dimensions Automotive: 980 mm × 850 mm × 650 mm (L × W × H)
Airfield: 980 mm × 850 mm × 650 mm (L × W × H)
1 Depending on the measuring task and the front lens, it is possible to measure larger objects.
2 CIE244:2021 f25

Why KMP Is the Smarter Choice for Light Distribution Measurement

  • Compact installation fits limited spaces and production environments
  • High-speed acquisition enables integration into production / EOL workflows
  • Full photometric outputs (luminous intensity, illuminance, color data) in a single system
  • Tailored calibration & configuration ensures optimal performance for your measurement
  • Scalable solution: move from R&D to production without changing measurement principles

One System for Every Lighting Task

KMP is suited for:

  • Automotive headlamp and DRL module testing
  • General lighting and luminaire measurement
  • Inline or end-of-line testing setups
  • Lighting laboratories where footprint is limited

Get Expert Guidance for Your Measurement Setup

Ready to speed up your light distribution measurements? Our specialists will help you configure the ideal KMP system for your specific application — from automotive headlamps to general lighting and EOL testing.
Contact TechnoTeam today for a personal consultation or a detailed quotation tailored to your requirements.

 
Type:
Package
Applications:
Automotive Aviation Light sources & luminaires
Measurands:
Light measurement
Tasks:
Automation & Industry